Inflow Velocity Calculation - Secondary Method

If the primary method cannot be performed, the secondary inflow calculation method

should be used. The secondary method uses a hot wire thermal anemometer. In this

method, the sash is lowered, and a series of velocity readings are taken at the front of

the unit.
NOTE: You must use Labconco Certifier Kit, Labconco P/N 3858403 to

perform this test properly. Qualified certifiers can obtain this kit by

contacting Labconco Customer Service Department at 800-821-5525.

For each model Purifier, the sash should be closed until the appropriate marked

template contacts the sash stop and the bottom edge of the sash, as shown below in

Figure 5-10.

The thermal anemometer probe should be positioned such that the distance from the

bottom of the probe holder to the center of the sensor element equals “Sensor Distance”

as shown in Table 5-5. Use the marks on the 6-foot templates, or a scale as shown in

Figure 5-11 & 5-12 below.

Figure 5-10


Measuring the restricted access opening

height, using the appropriate template.

Ensure that the bottom of the template is
seated in the sash stop.