Electrical Requirements
The biosafety cabinet models have the following electrical requirements:
Model #
Requirements
30441xx0x, 30448xx0x
115 VAC, 60 Hz, 16 Amps
30441xx1x, 30448xx1x
100 VAC, 50/60 Hz, 16 Amps
30441xx-20, 30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
30448xx-20, 30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
30461xx0x, 30468xx0x
115 VAC, 60 Hz, 16 Amps
30461xx1x, 30468xx1x
100 VAC, 50/60 Hz, 16 Amps
30461xx-20, 30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
30468xx-20, 30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
Note: A dedicated outlet with an appropriate circuit breaker should
be located as close as possible to the right rear side of the cabinet,
at a height even with, or higher than, the top of the bench or stand.
Consult your local electrical codes for properly rated circuit
breakers. For safe operation the dedicated outlet must provide the
protective earthing ground connection to the cabinet.
Note: On 100 and 115 VAC models, both electrical outlets are protected by
a ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.
Electrical outlets in the cabinet are restricted to 5 amps maximum
current.
Prises électriques dans l'armoire sont limitées à 5 courant maximum
ampères.
The biosafety cabinet models have the following electrical requirements:
Do not use any detachable power cord that is not adequately rated
for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
Table 2-1