Electrical requirements – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 42

Advertising
background image

Product Service 1-800-522-7658

42

Electrical Requirements

The biosafety cabinet models have the following electrical requirements:

Model #

Requirements

30441xx0x, 30448xx0x

115 VAC, 60 Hz, 16 Amps

30441xx1x, 30448xx1x

100 VAC, 50/60 Hz, 16 Amps

30441xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30448xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30461xx0x, 30468xx0x

115 VAC, 60 Hz, 16 Amps

30461xx1x, 30468xx1x

100 VAC, 50/60 Hz, 16 Amps

30461xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30468xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

Note: A dedicated outlet with an appropriate circuit breaker should

be located as close as possible to the right rear side of the cabinet,

at a height even with, or higher than, the top of the bench or stand.

Consult your local electrical codes for properly rated circuit

breakers. For safe operation the dedicated outlet must provide the

protective earthing ground connection to the cabinet.
Note:
On 100 and 115 VAC models, both electrical outlets are protected by

a ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.


Electrical outlets in the cabinet are restricted to 5 amps maximum

current.

Prises électriques dans l'armoire sont limitées à 5 courant maximum
ampères.
The biosafety cabinet models have the following electrical requirements:

Do not use any detachable power cord that is not adequately rated

for the unit.

Ne pas utliser un fil électrique amovible qui n’est pas du tension

nominale de l’appareil.

Table 2-1