Preparing the axiom for gaseous sterilization – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 112

Advertising
background image

Product Service 1-800-522-7658

112

Preparing the Axiom for Gaseous Sterilization

Note: This section only reviews the steps required for preparing the Axiom for

Gas Sterilization. Thoroughly understand the sterilization procedures and

protocols supplied by the manufacturer of the sterilizing system before

attempting this operation.

1.

Unplug the Axiom from electrical power.

2.

Thoroughly surface decontaminate the working area of the cabinet.

In order to get a complete seal on the front of the cabinet, do the following:

3.

Remove the work surfaces after it has been surface decontaminated.

4.

Remove the front grille.

5.

Place the work surfaces and the grille into the work area for

decontamination.

6.

Remove the front dress panel.

7.

Remove the sash stop bracket, located near the top of the blower plenum

cover.

8.

Lift the sash up until the bottom edge of the sash clears the bottom edge of

the Supply HEPA Filter shelf.

9.

Using flexible plastic sheet 5-8 mil thick and duct tape seal the perimeter of

the front opening.

10.

Using flexible plastic sheet and duct tape seal the Exhaust filter opening.

11.

The Axiom is now prepared for gaseous decontamination.