HEPA Leak Test Information

All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Axiom have been appropriately

decontaminated before proceeding.

Supply HEPA Filter Leak Test

All Axiom models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Axiom have been appropriately

decontaminated before proceeding.

Note: For all Axiom models, the work surfaces may be removed if desired, and

the aerosol generator discharge should be placed to the left of (but NOT in)

the center exhaust duct nearby to inject the aerosol as shown in Figure 5-1.

Figure 5-1

Place aerosol generator discharge into
the left recirculation duct


Center exhaust duct