Downflow Velocities

All Axiom models, are classified as having uniform downflow, and the average downflow

velocity for all 4-foot models is 65+/-5 FPM, and all 6-foot models is 55+/-5 FPM, tested

as per ANSI/NSF Standard 49:2012.

All readings should be in a plane 4 inches above the bottom edge of the sash, and the

UV Lamp, IV bar and all other accessories must be removed before performing

downflow test. The test grid dimensions are outlined in Table 5-3. When taking readings

in the back two rows, the anemometer probe should be pointing toward the nearest side

wall, as shown in figure 5-8. When reading the front row, the probe should point

outward, toward the front of the cabinet, as shown in Figure 5-9. The thermal

anemometer’s time constant should be set at 10-15 seconds for greater readout

stability.

Table 5-3

Downflow Test Specifications

Cabinet Width (Feet)

4

4

6

6

Sash Height (inches)

8

10

8

10

Nominal Avg. Downflow

(FPM)

65+/-5

65+/-5

55+/-5

55+/-5

Downflow test grid # of

points

24

24

36

36

Test Grid (# rows x #

columns)

3 x 8

3 x 8

3 x 12

3 x 12

Test Grid Data (in

inches)

Grid distance from back & sides

6.0

6.0

6.0

6.0

Distance between rows

5.75

5.55

5.75

5.55

Distance between columns

5.21

5.21

5.50

5.50