Optional ultraviolet lamp, Lamp socket replacement, Lamp identification – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 114: Uv lamp specifications

Advertising
background image

Product Service 1-800-522-7658

114

Lamp Socket Replacement

1.

Unplug the Axiom

2.

Remove the dress panel as noted in Figure 6-1

3.

Remove the fluorescent lamps from the sockets.

4.

To replace a socket, insert a small straight bladed screwdriver into the socket

near the lamp wire, as shown in Figure 7-1, and pull the wire out.

5.

Press the wire into the new socket.

6.

Reassemble the unit.

Figure 7-1


Optional Ultraviolet Lamp

Lamp Identification

All Axioms use a single 30-Watt germicidal lamp. The lamp information is shown

below. The code of the lamp part number is as follows:

G30T8

G = germicidal lamp

30 = the lamp Wattage

T8 = the lamp mounting configuration

UV lamp Specifications

For all models, the UV lamp number is G30T8. The Labconco part number

is 1271300.

For all 115/230 VAC models, the ballast assembly is Labconco part

number 3829901. The ballast is Robertson part number

PSM2GPH18MVW.

For all 100 VAC models, the ballast assembly is Labconco part number

3830600. The ballast is Robertson part number RSO1GPH30100.