Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 148

Advertising
background image

Product Service 1-800-522-7658

148


Relay Contact #2……the “airflow error” contact

This set of contacts (dry contacts pins 1 & 2 of J3 on the electrical enclosure, and the

ORANGE wires on Labconco harness P/N 3432300) will indicate when the Axiom has

detected an airflow error, as described by the conditions below:

Type of Axiom
unit

Alarm
Condition

LCD display
message

Airflow Error
dry contacts

Comments

All Labconco
Axiom Cabinets

Sash out of
position

SASH TOO HIGH

No Response

Motor RPM has

shifted abruptly

AIRFLOW ALERT

Contacts shorted

HEPA Filter
approaching end
of life—20%

CHECK FILTER
(Filter life indicator
at 20% or less)

No response

HEPA Filter

approaching end
of life-10%

CHECK FILTER

(Filter life indicator
at 10% or less)

Contacts shorted

HEPA Filter end

of life- 0%

CHECK FILTER

(Filter life indicator
at 0%)

Contacts shorted

Unit beeps

approximately
every 15 seconds

**Note all
specialty Axiom
BSC’s below will

have all the
alarm conditions

listed above, plus
specific alarms
listed below”

A2 BSC with
canopy sensor**

( see above)

Canopy Airflow
incorrect

CANOPY ERROR

Contacts shorted

A2 BSC with

airflow sensor**(
see above)

Downflow Airflow

incorrect

Alarm - Airflow

Contacts shorted

B2 BSC**( see
above)

Exhaust airflow
not correct

Alarm - Inflow

Contacts shorted