Product Service 1-800-522-7658

128

5.

Gently remove the Cable Sealing grommet, located on the side

of the exhaust duct. Remove the solid square plug. Locate the

replacement plug in the air sensor kit, and push the sensor

wiring harness through the slit in the plug.

NOTE: The grommet and plugs are keyed – they only fit

together properly when oriented as shown in the illustration.

Ensure both the replacement plug is oriented properly in the

grommet before installation.

Route the sensor cable from the sensor, through the grommet,

and outside of the exhaust duct. Reinstall the grommet. Ensure

the cable does not hinder the operation of the front relief valve.