Advertising
background image

Product Service 1-800-522-7658

56

3.

The screen will now prompt you for the password;

it is:

Light

UV Light

Timer

Timer

OK


Note: Any other key sequence will return you to the Attention screen.

4.

The first screen allows you to select whether the

Axiom is connected to an exhaust system or not. IT
IS CRITICAL THAT THE

THIS CABINET HAS

NO EXHAUST CONNECTION

OPTION IS

SELECTED. Press

OK

to continue.



.

5.

The second screen allows you to set the sash height

at either 8 or 10 inches. IT IS CRITICAL THAT

THE YOU LEAVE THE SASH HEIGHT AS IT WAS

SET AT THE FACTORY, UNLESS YOU WISH TO

RECERTIFY THE CABINET AT ITS NEW

SETTING.

6.

The third screen allows you to select whether the

Axiom has a UV light or not. IT IS CRITICAL

THAT THE YOU LEAVE THE UV LIGHT

CONFIGURATION AS IT WAS SET AT THE

FACTORY. Press

OK

to return to the first

Configuration screen.


The unit is now properly configured for operation recirculating its filtered exhaust air

into the room. In this mode, the unit will NOT display an exhaust alarm with the inlet

relief valve closed during operation.
NOTE: If you ever want to reconnect the Axiom to the exhaust system, you must

reconfigure it as such.