Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 56

Product Service 1-800-522-7658
56
3.
The screen will now prompt you for the password;
it is:
Light
UV Light
Timer
Timer
OK
Note: Any other key sequence will return you to the Attention screen.
4.
The first screen allows you to select whether the
Axiom is connected to an exhaust system or not. IT
IS CRITICAL THAT THE
THIS CABINET HAS
NO EXHAUST CONNECTION
OPTION IS
SELECTED. Press
OK
to continue.
.
5.
The second screen allows you to set the sash height
at either 8 or 10 inches. IT IS CRITICAL THAT
THE YOU LEAVE THE SASH HEIGHT AS IT WAS
SET AT THE FACTORY, UNLESS YOU WISH TO
RECERTIFY THE CABINET AT ITS NEW
SETTING.
6.
The third screen allows you to select whether the
Axiom has a UV light or not. IT IS CRITICAL
THAT THE YOU LEAVE THE UV LIGHT
CONFIGURATION AS IT WAS SET AT THE
FACTORY. Press
OK
to return to the first
Configuration screen.
The unit is now properly configured for operation recirculating its filtered exhaust air
into the room. In this mode, the unit will NOT display an exhaust alarm with the inlet
relief valve closed during operation.
NOTE: If you ever want to reconnect the Axiom to the exhaust system, you must
reconfigure it as such.