Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 131

Advertising
background image

Product Service 1-800-522-7658

131

Service Fixture Kit No. 3747500

Completely decontaminate all interior work surfaces of the Purifier before

beginning the installation of the service valve.

Additional service fixtures can be installed in any of the four available service fixture

holes in the sidewalls. The service fixture kit is Labconco part number 3747500.

Note: Some models have a solenoid valve connected to the service valve on the right

side, rear position. The solenoid prevents gas from flowing to the service valve when the

unit blower is off. It is the only service valve position that can be fitted with a solenoid

valve. Connect the gas service to the solenoid valve.

The service lines (if any) should be connected to the tube fitting(s) on the outside of the

liner wall as shown below. To install the tubing, follow these steps:
1.

Decide which of the available locations you want to install the service fixture.

2.

Access the exterior surface of the sidewall by removing the two Phillips screws on

the front edge of the panel. Swing open and lift off the panel to access the valve

hole plug.

3.

With a flat blade screwdriver, carefully break the silicone seal on the perimeter of

the hole plug (on the outside of the cabinet). Then pry the plug away from the side

of the cabinet.

4.

Ensure that the tubing is ¼ inch O.D., soft metal, and that the end has been

completely deburred.

5.

Route the tubing from the rear of the cabinet, ensuring that it will line up with the

slot in the back of the side panel. The slot is located from 8 ¾ to 11 ¼ inches (222

to 288 mm) from the bottom of the cabinet.

Note: Make sure that the tube routing will not contact any electrical wires.

DO NOT loop service line tubing within the side panels of the cabinet.

6.

Make sure that the nut on the tube fitting is loose, but do not remove it. Look inside

the fitting to make sure the tube ferrule is there.

7.

Push the tube into the fitting until it is properly seated. The tube will go

approximately ¾ inch (19 mm) into the fitting.

8.

Tighten the tube fitting nut hand tight and then, using a 7/16-inch wrench, tighten it

at least ¾ turn more.

9.

Close the service valve in the biosafety cabinet and then slowly open the shutoff

valve on the service valve. Test all fittings for leakage. Tighten the tube nut slightly

if needed.

Advertising