Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 140

Advertising
background image

Product Service 1-800-522-7658

140


4)

Locate the internal RS-232 cable (P/N 3432000) included in the kit and connect it

between the RS-232 PCB J1 and the Transition PCB (P/N 3431500) J6 as shown

in picture 5.

Picture 5

5)

Re-secure the electrical control box to the top of the Axiom.

6)

Connect the computer interface cable (P/N 7537800 – 9 pin or P/N 7537801 – 25

pin) included in the kit to the RS-232 PCB J2 connector, and connect the other

end of this cable to a 9-pin or 25 pin serial input of a computer. Proceed to the
directions “Using the RS-232 Receptacle”.

7)

To enable access to the Alarm/Blower contacts connect the alarm/contactor

cable (P/N 3432300) included in the kit to the RS-232 PCB J3 connector. J3 is

shown above the “ALARM CONTACT” label in picture 1.

Signal

Wire Color

J3

ALARM 1 – Airflow

ORANGE

1

ALARM 1 - Airflow

ORANGE

2

ALARM 2 - Blower

WHITE

3

ALARM 2 - Blower

WHITE

4

GROUND

BLACK

5

+12V

BLUE

8