When requested for the password press

Light

,

UV light

,

Timer

,

Timer

then

OK/Mute

.





16.

If the password is properly entered, the first certification screen

will display the supply blower motor’s PWM setting. Press the

or

button to increase or decrease the blower speed as

needed. When the speed is set properly, press the

OK/Mute

button to lock the new blower speed setting, and continue to the
filter life screen.


17.

If this is an initial certification or a recertification after supply

HEPA filter replacement, select

Reset Filter Gauge to 100%

. If

this is an annual recertification, select

Leave Gauge unchanged

.

If you want to set the filter gauge at a preset value, select

Set

Gauge to new setting

. Then press

OK/Mute

.



18.

The next certification screen will display the exhaust blower

motor’s PWM setting. Press the

or

button to increase or

decrease the blower speed as needed. When the speed is set
properly, press the

OK/Mute

button to lock the new blower

speed setting, and continue to the filter life screen.




19.

If this is an initial certification or a recertification after exhaust

HEPA filter replacement, select

Reset Filter Gauge to 100%

. If

this is an annual recertification, select

Leave Gauge unchanged

.

If you want to set the filter gauge at a preset value, select

Set

Gauge to new setting

. Then press

OK/Mute

.





20.

If the unit is has no airflow sensor, after the filter gauge option

is selected and you press

OK/Mute

, the Axiom will power

down, and then restart with the new settings. If the Unit has an
airflow sensor, you will then be asked to calibrate the inflow

sensor to match the value you obtained when you measured the
inflow. After the inflow is calibrated, and you press

OK/Mute

,

the Axiom will power down, and then restart with the new
settings.

PASSWORD:

Calibrat ion

The Inflow displayed

value is: 102 FPM

Calibrat ion

Inf low Display

or

t o change display value,

UV Light t o reset t o 105 FPM,

OK t o select , MENU t o f inish

 

The ECM set t ing is: 80

The current RPM is: 1050

The average RPM is: 1020

Calibration

Supply Fan Speed

or

t o change mot or speed,

OK t o select , MENU t o escape

 

Calibration

Supply Filter Gauge

or

t o change f ilt er gauge,

OK t o select , MENU t o escape

 

Reset Filt er Gauge t o 100%

Leave Gauge unchanged

Set Gauge t o new set t ing

The ECM set t ing is: 80

The current RPM is: 1050

The average RPM is: 1020

Calibration

Exhaust Fan Speed

or

t o change mot or speed,

OK t o select , MENU t o escape

 

Calibration

Exhaust Filter Gauge

or

t o change f ilt er gauge,

OK t o select , MENU t o escape

 

Reset Filt er Gauge t o 100%

Leave Gauge unchanged

Set Gauge t o new set t ing