Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 135

Advertising
background image

Product Service 1-800-522-7658

135

Note: Any other key sequence will return you to the Attention screen.

9.

The first screen allows you to select whether the

Axiom is connected to an exhaust system or not.

IT IS CRITICAL THAT THE CABINET

MATCHES THE OPTION THAT IS SELECTED.

Press

OK

to continue.

10.

If connected to an exhaust system, the second

screen allows you to set how long the blowers in
the Axiom will continue to operate in the event of

an exhaust system failure. The interval can be

programmed from 0-300 seconds.



Note: Consult with your facility’s safety officer or Labconco to help establish how long

the Axiom should continue to operate after an exhaust system failure.

11.

The third screen allows you to set the sash height

at either 8 or 10 inches. IT IS CRITICAL THAT

THE YOU LEAVE THE SASH HEIGHT AS IT
WAS SET AT THE FACTORY, UNLESS YOU

WISH TO RECERTIFY THE CABINET AT ITS

NEW SETTING.

12.

The fourth screen allows you to select whether the

Axiom has a UV light or not. IT IS CRITICAL

THAT THE YOU SELECT

A UV LIGHT

. Press

OK

to return to the first Configuration screen.

The unit is now properly configured for operation with the UV light.

Refer to the user’s manual for further information regarding the proper

operation and maintenance of the UV light.