QuickChart – Axiom Type C1 BSCs – Imperial Units

Model

30441

30448

30461

30468

Type

C1

C1

C1

C1

Cabinet Size (in feet)

4

4

6

6

Sash Opening (inches)

10

8

10

8

Starting Serial #

1503_

1503_

1503_

1503_

Downflow Data

Nominal Avg. Downflow (FPM)

65+/-5

65+/-5

55+/-5

55+/-5

grid # of points (rows x columns)

24 (3x8)

24 (3x8)

36(3x12)

36(3x12)

Grid distance from back & sides

6.0

6.0

6.0

6.0

Distance between rows

5.55

5.75

5.55

5.75

Distance between columns

5.21

5.21

5.50

5.50

Inflow Data

Nominal Average Inflow (FPM)

105+/-5

105+/-5

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

3.37

2.69

5.03

4.03

Nominal Avg. Exhaust Vol. (CFM)

354

283

529

423

Avg. Exhaust Vol. Range (CFM)

337-371

269-296

503-553

403-443

Secondary Inflow Data

Sash Opening Template

1

4+

4+

6+

6+

Sensor distance (inches)

2

3.25

3.25

4

4

# of Test points

8

8

12

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

354

283

289

231

Avg. Inflow Vel. Range (FPM)

312-344

249-274

252-277

202-222

Correction Factor (CF)

1.08

1.08

2

2

Avg. Inflow Volume (CFM)

337-371

269-296

503-553

403-443

Sash Open Area (Sq. Ft)

3.37

2.69

5.03

4.03

Inflow Velocity Range (FPM)

100-110

100-110

100-110

100-110

Supply HEPA Data

Air Displacement (CFM)

530

530

660

660

Laskin Nozzles needed

1

1

1

1

Theoretical aerosol conc. (ug/l)

3

25

25

20

20

Width x Depth x Height (in.)

4

48x18x3.06

48x18x3.06

72x18x3.06

72x18x3.06

Performance (CFM)

445

445

665

665

Performance (Pressure in. H

2

O)

<.45”

<.45”

<.42”

<.42”

Labconco P/N

3838401

3838401

3838403

3838403

Exhaust HEPA Data

Air Displacement (CFM)

354

283

529

423

Laskin Nozzles needed

1

1

1

1

Theoretical aerosol conc. (ug/l)

3

38

48

26

32

Width x Depth x Height (in.)

4

26x18x5.88

26x18x5.88

48x18x5.88

48x18x5.88

Performance (CFM)

354

354

529

52

Performance (Pressure in. H

2

O)

<0.3

<0.3

<0.3

<0.3

Labconco P/N

3838501

3838501

3838503

3838503

Exhaust Volume Data

DIM Exhaust Volume (CFM)

5

400

323

570

463

Traverse Exhaust Volume (CFM)

6

480

387

684

556

Differential Pressure (in. H

2

0)

7

0.3

0.3

0.3

0.3

IRV Inflow Range (CFM)

8

20-100

20-100

20-100

20-100

Nominal IRV Slot Velocity (FPM)

240-260

240-260

240-260

240-260

Nominal IRV Slot Volume (CFM)

50

50

50

50

Supply Motor/Blower Data

Motor HP

1/3

1/3

1/3

1/3

Exhaust Motor/Blower Data

Motor HP

1/3

1/3

1/3

1/3

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F32T8 TL741

F32T8 TL741

F40T8 SP41

F40T8 SP41

Fluor. Lamp Labconco Part #

9721900

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100