Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 78

Advertising
background image

Product Service 1-800-522-7658

78

3.

Pull the tube toward the front of the unit. Remove the access tube cover.

4.

Using an appropriate photometer, connect the unit's upstream line to the access

tube.

Note: Keep the photometer sampling tube as short as possible. It should

never exceed 6 feet maximum.

1.

Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2.

Check the oil level of the generator – it must be within 1/8” of the level line

when the generator is sitting level.

3.

Establish the 100% and 0% concentration levels for the photometer.

4.

Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

5.

Remove the diffuser by raising the sash as high as it will go, and then removing

the two acorn nuts with a 7/16” wrench or socket, as shown in Figure 5-3.

Figure 5-3

Removing the diffuser acorn nuts (Dress panel and

lamps removed for clarity only).

Note: The diffuser acorn nuts are secured to the studs with a removable thread locking

compound to prevent them from vibrating loose during shipment of the Axiom. There

may be some resistance the first time the nuts are removed; this is normal. No

additional thread locking compound need be applied to these acorn nuts unless the unit

is being prepared for shipment.

1.

Lower the front edge of the diffuser until it clears the bottom of the sash, and

then pull the diffuser straight out of the cabinet, as shown in Figure 5-4.

Figure 5-4


Advertising