Quickstart


How do I run a diagnostic on the Inlet Relief Valve?


1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the

Diagnostic

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to enter the Diagnostic submenu.






6.

Press the

button until the Diagnostic

Inlet Relief Valve

option

is highlighted (it will turn white). Press

OK/Mute

.






7.

The display shows the status of the Inlet Relief Valve. During

normal operation, it will be open, and during an exhaust alarm,
it will be closed. NOTE: The inlet relief valve function is
OPPOSITE OF THE MAGNETIC SWITCH FUNCTION! – WHEN

THE INLET VALVE IS OPEN, THE MAGNETIC SWITCH IS
CLOSED.








MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Calibration

Configuration

Diagnostics

 

Service

Diagnost ics

or

t o choose,

OK t o select , MENU t o escape

Sash sensors

Keypad - Relays

Inlet Relief Valve

Airf low Sensor

 

Diagnost ics

Test t he alarm by closing t he

inlet valve.

Note the alarm st at us below:

Open

OK t o ret urn