Quickstart
How do I run a diagnostic on the Inlet Relief Valve?
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Diagnostic
option is highlighted (it
will turn white).
5.
Press
OK/Mute
to enter the Diagnostic submenu.
6.
Press the
▼
button until the Diagnostic
Inlet Relief Valve
option
is highlighted (it will turn white). Press
OK/Mute
.
7.
The display shows the status of the Inlet Relief Valve. During
normal operation, it will be open, and during an exhaust alarm,
it will be closed. NOTE: The inlet relief valve function is
OPPOSITE OF THE MAGNETIC SWITCH FUNCTION! – WHEN
THE INLET VALVE IS OPEN, THE MAGNETIC SWITCH IS
CLOSED.
MAIN MENU
1. MyLogic
TM
2. Set t ings
3. Service
or
t o choose,
OK t o select , MENU t o escape
or
t o choose,
OK t o select , MENU t o ret urn
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service
Diagnost ics
or
t o choose,
OK t o select , MENU t o escape
Sash sensors
Keypad - Relays
Inlet Relief Valve
Airf low Sensor
Diagnost ics
Test t he alarm by closing t he
inlet valve.
Note the alarm st at us below:
Open
OK t o ret urn