Concurrent Balance Values

Axiom Inflow and Inlet Valve flow numbers are established by a direct measurement of

air flow. Traditional exhaust system measurements are made using either an

anemometer or Pitot traverse methodology. Because of these different methodologies,

the total volume of air flowing through the exhaust system will vary.

Table 5-6

Concurrent Balance Volume Specifications

Model

30448

30441

30468

30461

DIM-measured Inflow (CFM)

283

354

423

529

Nominal Inflow through Valve (CFM)

40

40

40

40

Total DIM-measured (CFM)

323

394

473

569

Traverse-measured Exhaust (CFM)

387

480

556

684