Section Nine –

Miscellaneous

Troubleshooting

Refer to the following table if the Biosafety Cabinet fails to operate properly. If the

suggested corrective actions do not solve the problem, contact Labconco for additional

assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower and
lights won’t turn on

Unit not plugged
into outlet

Plug the Biosafety Cabinet into
appropriate electrical service.

System Reset

Switch is Off

Turn on the System Reset Switch.

Circuit breaker(s)

tripped

Reset circuit breakers.

Keypad

disconnected or

defective

Run switch diagnostics and check

connections.

Sash closed-unit is
in Night-Smart

Mode

Raise sash.

Sash Closure

Switch is

disconnected or

defective

Run switch diagnostics and check

connections.

Blower does not run

or only runs at low

speed, but lights

work

Sash Restricted

opening Switch is

disconnected or

defective

Run switch diagnostics and check

connections.

Motor signal wires

damaged or

disconnected

Measure motor output signal

voltage. Inspect signal wiring and

connections

Power Supply

Board

misconnected or

defective

Test output voltages on the Power

Supply board.