Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 79

Advertising
background image

Product Service 1-800-522-7658

79

2.

The Axiom is now ready for the Supply HEPA Filter Leak Test. Set the

photometer sampling valve to “DOWNSTREAM”. Ensure there is a proper vacuum

at the sampling port of the pistol.

3.

Open one Laskin nozzle.

4.

Ensure the pressure gauge on the generator is reading 22 +/- 1 PSIG.

5.

Scan the downstream side of the HEPA filter by passing the sampling nozzle of

the gun in slightly overlapping strokes over the entire surface of both of the

filters, with the sampling port not more than 1 inch from the surface of the filter

media. Scan the entire periphery of the filters and the gaskets between the filter

frame and the Axiom frame. Scanning shall be done at a traverse rate of not

more than 2 inches per second.

Figure 5-5

Note: Operation of the photometer may become erratic
when testing the front edge of the supply filter frame,

due to aspiration of room air into the front of the work

area. This problem can be minimized or eliminated by

placing the edge of a sheet of rigid plastic or metal at

the edge of the filter gasket when scanning this area, as

shown in Figure 5-15.

Acceptance
Aerosol penetration shall not exceed 0.01%.

Reassembly

When reinstalling the supply filter diffuser, ensure that the two holes in the back edge of

the diffuser engage the diffuser support posts on the rear wall of the work area, as

shown in Figure 5-6. Reinstall the prefilter.

Figure 5-6

Diffuser pins engaging the holes in the rear of

the diffuser.