Timer operation, Interval timer operation, Stopwatch timer operation – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 72: If an airflow alert activates, Resetting the airflow alert system

Advertising
background image

Product Service 1-800-522-7658

72

Timer Operation

NOTE: The timer button allows activation of an interval (countdown) or

stopwatch (elapsed) timer. The timers cannot be operated simultaneously.

To access the main timer menu, press the

Timer

button anytime during normal

operation. The main timer menu is shown on the LCD display. Use the

and

buttons

to highlight the

Interval

or

Stopwatch

Timer. Press the

OK/MUTE

button to select the

highlighted timer function.

Interval Timer Operation

1.

When selected, the Interval Timer menu is displayed on the LCD. The timer

defaults to 0:00:00 (hours:minutes:seconds).

2.

Press and hold the

or

buttons to increase or decrease the timer interval.

3.

When the proper interval is entered on the display, press the

OK/MUTE

button

to start the timer.

4.

When the timer reaches 0:00:00, an audible alarm will sound, and the timer will

reset itself and repeat the countdown.

5.

Press the

OK/MUTE

button to pause the timer.

6.

Press the

Menu

button to clear the interval timer and return to the main timer

menu.

Stopwatch Timer Operation

1.

When selected, the Stopwatch Timer menu is displayed on the LCD. The timer

defaults to 0:00:00.

2.

Press the

OK/MUTE

button to start the timer.

3.

Press the

OK/MUTE

button again to zero the timer.

4.

Press the

Menu

button to clear the stopwatch timer and return to the main

timer menu.

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:

Blockage of the inlet grilles or exhaust outlet.

Removal of the work surface or grille during operation.

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation once the motor speed has

stabilized.