5 dos u, 1 dos u - overview, Dos u – Metrohm tiamo 2.1 Manual User Manual

Page 1075

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

1061

[Delete]

Deletes the calculated measured value selected in the table.

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can be
taken over and saved from existing measuring commands (i.e. MEAS) which
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 858)
.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 858)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.6.3.5

DOS U

5.6.6.3.5.1

DOS U - Overview

Dialog window: Method

DOS U Properties... DOS U - 'Command name'

Command for controlled dosing with measured value U.

Devices
This command can be executed with the following devices:

Titrando: 835, 836, 842, 857, 902, 906, 907

Titrino: 718, 736, 751, 799

pH/Ion Meter: 867

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Advertising