5 dosing - empty, 6 dosing - add fix volume, Empty – Metrohm tiamo 2.1 Manual User Manual

Page 1435: Fixed volume

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual Control

tiamo 2.1

■■■■■■■■

1421

8.4.1.5

Dosing - Empty

Program part: Manual Control

Here the emptying of dosing units can be started and stopped. This tab
appears when selecting a single dosing device as well as with all dosing
devices with dosing units.

Note

This tab is not displayed for external dosing devices on the Titrinos 736,
751, 758, 799
.

[Start]

Start emptying for the selected dosing device(s). The parameters defined for
the dosing units will be used.

[Stop]

Stop emptying for the selected dosing device(s).

8.4.1.6

Dosing - Add fix volume

Program part: Manual Control

Here the dosing of a preset volume can be started and stopped. This tab
only appears if a single dosing device is selected.

Volume

Fixed volume to be dosed.

Range

100 ... 99999.9 ml

Default value

100 ml

Dosing rate

The volume is dosed at this rate. The maximum dosing rate depends on the
cylinder volume of the exchange unit or dosing unit used. If the entered
dosing rate is too high for the selected dosing device, it will automatically
be reduced during dosing to the largest possible value.

Note

For viscous liquids the dosing rate should be reduced.

Selection

maximum

Default value

maximum

Titrando, 814, 815, 855
Range

0.01 ... 166.00 mL/min

Advertising