5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

722

■■■■■■■■

tiamo 2.1

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the Event
message
Start titration is sent to the command with a SEND command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.

5.6.3.4.6.8

SET Upol - Additional evaluations

Tab: Method

SET Upol Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value
, time or volume) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).