3 met, 1 met - overview – Metrohm tiamo 2.1 Manual User Manual

Page 549

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

535

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.3

MET

5.6.3.3.1

MET - Overview

Menu item: Method

Paste New command...

Command for Monotonic Equivalence Point Titrations.

Principle
With this command titrations with reagent addition at constant volume
increments are carried out. Measured value acceptance is drift-controlled
(equilibrium titration) or after a waiting time.

Note

This command is suited for titrations with relatively high signal fluctuations
or suddenly occurring potential jumps and for slow titrations or slowly
responding electrodes.

Evaluation
The equivalence points (EPs) are localized by a procedure based on the For-
tuin method which has been adapted by Metrohm for numerical methods.
A search is made for the largest measured value modification (

n

). The exact

EP is determined by using an interpolation factor

ρ which depends on the

values before and after

n

.

V

EP

= V

0

+

ρ ∆V