5 stat u - titration parameters, Titration parameters – Metrohm tiamo 2.1 Manual User Manual

Page 855

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

841

5.6.3.8.7.5

STAT U - Titration parameters

Tab: Method

STAT U Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

Parameters defining the run of the titration.

Titration direction

Selection of the titration direction. If two endpoints are set, then the titration
direction is already established and the following selection of titration
direction
is ignored.

Selection

+ | - | auto

Default value

auto

+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Measuring point recording

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Advertising