5 properties - data import, Properties - data import – Metrohm tiamo 2.1 Manual User Manual

Page 122

Advertising
background image

3.3 Sample tables

■■■■■■■■■■■■■■■■■■■■■■

108

■■■■■■■■

tiamo 2.1

Stop series ([Stop])
If this option is selected, then the series is stopped before the start of the
marked line (equivalent to [Stop]).

Start the Text editor with which the message can be entered and edited.

3.3.2.5

Properties - Data import

Dialog window:Workplace

Tools Sample

table

New… [Sample table] Properties… Properties - Sample

table 'Name'

Dialog window:Workplace

Run Determination series Prop-

erties Run window

Properties - Determination series

Enabling/disabling the import of external data for sample table fields.

Note

The successful conclusion of the data import during the run will be sig-
naled by an acoustic signal (beep).

Importing from balance
on | off (Default value: off)

If this option is enabled, then external data from the selected balance will
be imported into the opened sample table. Data can be imported into the
fields ID1…ID4, Sample size and Unit.

Note

In order to be able to import sample identifications from the balance, data
export must be set accordingly at the balance (Use designations ID1,
ID2, ID3, ID4).

Balance

Selection of the balance from which data is to be imported.

Selection

Device name

Selection

Always accept data | Accept data only for
active window

Default value

Always accept data

Advertising