background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

572

■■■■■■■■

tiamo 2.1

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

Pause

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.3.3.4

MET U - Titration parameters

Tab: Method

MET U Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

Parameters defining the run of the titration.

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance
and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.