5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

828

■■■■■■■■

tiamo 2.1

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 858)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.8.6.10

STAT pH - Fix endpoint evaluation

Dialog window: Method

STAT pH Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Properties] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-20.000 ... 20.000 pH

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.8.7

STAT U

5.6.3.8.7.1

STAT U - Overview

Dialog window: Method

STAT U Properties... STAT U - 'Command name'

Command for STAT titrations with measured value U.

Devices
This command can be executed with the following devices:

Titrando: 835, 836, 842, 857, 902, 906, 907

Titrino: 718, 736, 751, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance: