9 brc - additional measured values, Additional measured values – Metrohm tiamo 2.1 Manual User Manual

Page 811

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

797

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value
, time or bromine) for the fixed endpoint from the measuring point
list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.7.10, page 798).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.7.10, page 798).

[Delete]

Deletes the selected line.

5.6.3.7.9

BRC - Additional measured values

Tab: Method

BRC Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings:

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can be
calculated from existing measured values and saved, which can also be pre-
sented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.

Advertising