5 Method

tiamo 2.1

■■■■■■■■

853

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 858)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.8.7.10

STAT U - Fix endpoint evaluation

Dialog window: Method

STAT U Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Properties] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.9

Evaluation

5.6.3.9.1

pK value and Half neutralization potential

Parameters: Method

DET/MET Properties... Additional evaluations pK/HNP

evaluation

The pK value can be determined for pH titrations (DET and MET) and the
half neutralization potential can be determined for U titrations.

The activities of conjugated acid-base pairs are linked by the following equa-
tion (Henderson-Hasselbalch equation):

pH = pK

a

+ log (a

B

/a

A

)