6 stat u - stop conditions, Stop conditions – Metrohm tiamo 2.1 Manual User Manual

Page 857

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

843

5.6.3.8.7.6

STAT U - Stop conditions

Tab: Method

STAT U Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stops when the preset time has elapsed after the start conditions have been
fulfilled.

Range

0 ... 999999 s

Selection

off

Default value

off

since

Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):

Selection

Start | Control point first reached | Last dosing

Default value

Start

Start
The stop time starts after the start volume has been dosed and after the
following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.