6 meas u - fix endpoint evaluation, 4 meas ipol, 1 meas ipol - overview – Metrohm tiamo 2.1 Manual User Manual

Page 898: Meas ipol

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

884

■■■■■■■■

tiamo 2.1

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 858)
.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 858)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.4.3.6

MEAS U - Fix endpoint evaluation #

Dialog window: Method

MEAS U Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Property] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value (Titrando, 867, 855)
Range

-2000.0 ... 2000.0 mV

Measured value (Titrino)
Range

-2000 ... 2000 mV

Time
Range

0.0 ... 999999.9 s

5.6.4.4

MEAS Ipol

5.6.4.4.1

MEAS Ipol - Overview

Dialog window: Method

MEAS Ipol Properties... MEAS Ipol - 'Command name'

Command for voltametric measurements with selectable polarization
current.

Devices
This command can be executed with the following devices:

Advertising