tiamo 2.1

■■■■■■■■

971

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.CONC

Concentration of the solution used for the command

.DBL

Total duration for the processing of the command in s

.EME

Final measured value (measured value after processing of
the command) in the unit of the measured value

.ENP

Electrode zero point calculated from the calibration

.ETE

End temperature (temperature following processing of the
command) in °C

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.IME

Initial measured value (measured value before start condi-
tions are processed) in the unit of the measured value

.ITE

Initial temperature (temperature before start conditions are
processed) in °C

.MTE

Temperature measurement with sensor; 1 = on, 0 = off)

.RES

Calculated result of the standard addition in the selected
unit

.SLO

Electrode slope calculated from the calibration

.STY

Type of stop with which the command was stopped: 1 =
normal
;0 = manual or after error

.VAR

Variance of the calculated result of the standard addition in
the selected unit

.VOL

Dosed volume

5.6.4.14.2

STDADD man - General/Hardware

Tab: Method

STDADD man Properties... General/Hardware

Command name

Name of the command.

Input

25 characters