5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

428

■■■■■■■■

tiamo 2.1

MET
Command for Monotonic Equivalence Point Titrations.

(see Chapter 5.6.3.4.1, page 639)
Command for Set Endpoint Titrations.

KFT
Command for volumetric Karl Fischer Titrations.

KFC
Command for coulometric Karl Fischer Titrations with voltametric meas-
urement.

BRC
Command for coulometric titration for determining the bromine con-
sumption.

STAT
Command for STAT Titrations.

5.6.3.2

DET

5.6.3.2.1

DET - Overview

Menu item: Method

Paste New command...

Command for Dynamic Equivalence Point Titrations.

Principle
With this command titrations with dynamic incremental reagent addition
are carried out. The volume increments vary as a function of the slope of the
curve. An attempt is made to achieve constant measured value modifications
with each dosing. The optimal volume for dosing is determined from the
measured value modifications of the previous dosings. Measured value
acceptance is drift-controlled (equilibrium titration) or after a waiting time.

Note

As the reagent addition in DET depends on the measuring data, the titra-
tion curve should not vary too much from an S-shaped curve.

Advertising