5 Method

tiamo 2.1

■■■■■■■■

965

Device

Device name

Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start of
the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Avantes Spectrometer

Default value

Avantes Spectrometer

5.6.4.13.3

MEAS Opt - Measuring parameters

Tab: Method

MEAS Opt Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded

Range

200.0 ... 1100.0 nm

Default value

400.0 nm

Integration time

Integration time for absorption measurement.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.