6 kfc - stop conditions, 7 kfc - conditioning, Stop conditions – Metrohm tiamo 2.1 Manual User Manual

Page 793: Conditioning

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

779

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

756, 831
Range

1 ... 999999 s

Default value

2 s

851, 852
Range

0.1 ... 999999.0 s

Default value

2.0 s

5.6.3.6.6

KFC - Stop conditions

Tab: Method

KFC Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the titration are defined on this tab.

Stop time

Stops when the preset time has elapsed since the start of the titration. off
means no stop.

Range

1 ... 999999 s

Selection

off

Default value

off

5.6.3.6.7

KFC - Conditioning

Tab: Method

KFC Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Conditioning

on | off (Default value: on)

Enable/Disable the conditioning. If the conditioning is disabled, the following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the