6-24 IETG FlowHawk Manual

Note: Modifications made to the Advanced Parameters
should be performed only by or under the guidance of
trained IETG personnel. Therefore, IETG recommends
leaving the default settings (Auto) in place.

Modify the parameters as necessary, and then select OK once
complete.

DAC Gain Choose the method for managing the

amplification of the analog signal that returns to the sensor:

Auto Select this option to allow the monitor to

automatically optimize the amplification of the signal.

Manual Select this option to enter a fixed value

between -12db (lower amplification) and +12db (high
amplification) at which to amplify the signal.

Ping Amplitude Choose the method for managing the

amplification of the analog signal that is sent out, or
transmitted from, the sensor. Smaller pipes typically
require weaker (quieter) amplification; larger pipes
typically require stronger (louder) amplification.

Auto Select this option to allow the monitor to

automatically optimize the amplification of the signal.

Manual Select this option to enter the percentage of

full amplification at which to transmit the signal.
Enter a percentage between 10 and 100 percent, where
100 percent represents maximum amplification.

Pressure tab

Edit the following parameters corresponding to the pressure depth
sensor in the Peak Combo Sensor:

Note: Keep in mind that the Peak Combo SL device (i.e.,
the Slimline Peak Combo Sensor) does not have a pressure
component. Therefore, Profile does not include pressure
parameters for this device.