8-42 IETG FlowHawk Manual

Problem

Monitor independently discontinues logging data.

Possible

Causes

Battery pack may be dead or below minimum voltage
requirement (7.5 volts - internal), or external power
source may be down.

3.3-volt fuse on regulator may have blown.

Monitor memory may have been corrupted during
firmware upgrade, causing the monitor to record data
only during monitor communications.

Possible

Solutions

Replace battery pack.

Replace the 3.3-volt yellow fuse (when replaceable) on
the regulator or return the monitor to IETG for power
regulator replacement.

If the monitor is logging data only during monitor
communications following a firmware upgrade, contact
your IETG representative to clear the monitor memory
and reactivate the monitor.

Contact your regional IETG representative.

Problem

An I/O error message displays when communicating
with the monitor.

Possible

Causes

Communication quality might be marginal.

Antenna may be defective.

Processor board may be faulty.

Possible

Solutions

Re-attempt communication with the monitor.

Temporarily attach new antenna, and permanently
replace the existing antenna if the error message no
longer occurs.

Replace the monitor if defective.

Contact your regional IETG representative.