Configuration and Activation 6-27

Note: The velocity parameters corresponding to the
Peak/Slimline Peak Combo Sensor typically should not
require editing. However, if modifications are necessary,
IETG does not recommend making changes to these
parameters without proper training or assistance.

Note: When using a Peak Combo Sensor and Surface
Combo Sensor at the same monitoring point, ADS
recommends maintaining the same settings for all of the
parameters corresponding to both Peak Velocity (Velocity
tab on the Edit Peak Combo Parameters dialog) and
Surcharge Peak Velocity (Peak Velocity tab in the Edit
Surface Combo Parameters dialog) to ensure consistency
between the velocity readings.

Enable Select this checkbox to ensure the monitor begins
taking velocity readings at the designated interval once
activated.

Flow Type Select the option that best reflects the kind of
flow conditions that typically occur at the monitoring point
from the drop-down list:

Typical Flow exhibits a typical diurnal pattern. This is

the default setting.

Wet/Stagnant Flow remains in the pipe, but moves only

during wet weather events.

Usually Dry Pipe typically remains dry, except under

wet weather conditions.

Sensitivity Select the appropriate velocity sensor sensitivity
from the drop-down list. A sensitivity that is too high may
produce noise, causing interference in the signal and resulting
in spikes in the data. A sensitivity that is too low can produce
readings that are low in comparison with confirmation readings.
The default sensitivity is 10.