ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 348

Advertising
background image

Specifications

A-3

• The log rate determines the number of readings taken per day.

At a 15-minute sample rate, the monitor will record 96 readings
a day for a single entity. At a 2-minute sample rate, the
monitor will record 720 readings a day for a single entity.

• One entity reading consumes 8 bytes of monitor memory.

• The following table lists the standard number of entities

recorded for each sensor device. The date/time stamp for a
reading is equal to one reading. Therefore, each total below
represents the actual entities plus one date/time stamp. A
date/time stamp must be included for each monitoring point
associated with the monitor.

Peak

Combo
Sensor

Surface

Combo
Sensor

Ultrasonic

Depth

Sensor

Ultrasonic

Depth Sensor

with Pressure

Option

Slimline

Peak

Combo
Sensor

7

6

3

4

6

• Deduct 1 to 2 days from every 25 days from the total days of

storage to compensate for additional overhead in memory.

Following is the equation for calculating the total days of storage
available in the monitor memory:

Data Storage Available in the Monitor (in Bytes)

=

Total Days of

Storage in

the Monitor

Memory

Number of

Readings

Taken per

Day for

One entity

X

Memory

Consumed for

One Entity

Reading

(Date/Time

Stamp)

X

Number of

Entities Logged

by the Monitor

(based on the

selected
devices)

For example, recording data at a 2-minute sample rate using a Peak
Combo Sensor offers approximately 20 days of data storage.

851,968 bytes

=

21 days (20 when

accounting for

overhead)

720 readings

X

8 bytes

X 7 entities (including

1 for the date/ time

stamp)

Advertising