Maintenance and Troubleshooting 8-49

Problem

Range from the ultrasonic depth subsystem is too short
(but not zero).

Possible

Causes

Pipe height may be incorrect.

Electronic offset may be incorrect.

CS3 depth sensor may be dirty.

CS3 depth sensor may be faulty.

Possible

Solutions

Verify the pipe height.

Verify electronic offset and adjust if necessary.

Clean the sensor.

Replace the sensor.

Contact your regional IETG representative.

Problem

Raw ultrasonic depth data shows depths greater than
the pipe diameter.

Possible

Causes

Physical offset may be incorrect.

Pipe height may be incorrect.

Electronic offset may be incorrect.

Pipe may be surcharged.

Sensor may be dirty.

Sensor may be faulty.

Possible

Solutions

Verify the physical offset.

Verify the pipe height.

Verify electronic offset and adjust if necessary.

Clean the sensor.

Replace the sensor.

Contact your regional IETG representative.