8-50 IETG FlowHawk Manual

Problem

Depth data indicates a surcharged pipe, but the pipe is
free flowing.

Possible

Causes

Physical offset may be incorrect.

Electronic offset may be incorrect.

Pipe height may be incorrect.

Sensor may be dirty.

Sensor may be faulty.

Possible

Solutions

Verify the physical offset.

Verify the pipe height.

Verify electronic offset and adjust if necessary.

Clean the sensor.

Contact your regional IETG representative.

Replace the sensor.

Replace the monitor if defective.

Problem

Ultrasonic depth readings are erratic.

Possible

Causes

Noise may exist in the sewer pipe.

Sensor may be dirty.

Flow may be choppy or foamy.

Sensor may not be level.

Possible

Solutions

Check the flow conditions.

Clean the sensor.

Check if the sensor is level and correct if necessary.