Pressure depth subsystem – ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 344

Advertising
background image

8-60 IETG FlowHawk Manual

Pressure Depth Subsystem

The following tables contain general troubleshooting techniques for
the pressure depth subsystem corresponding to the Peak Combo
Sensor, Surface Combo Sensor, and Ultrasonic Depth Sensor with
the pressure depth option.

Note: Contact your regional IETG representative for
further diagnosis before replacing a sensor.

Problem

Pressure depth readings are erratic.

Possible

Causes

Pressure vent tube may be pinched or blocked with
water.

Pressure dryer tube may not be functioning properly.

Black cap may still be on the dryer tube.

Possible

Solutions

Inspect color of desiccant in dry tube and replace the
dryer tube if necessary.

Inspect the vent tube along its entire length to ensure
the cable is not pinched and that the cable ties securing
the vent tube are not too tight.

Verify that the black cap has been removed from the
dryer tube.

Problem

Pressure depth readings display a good pattern, but are
consistently incorrect.

Possible

Cause

Sensor physical offset may be incorrect.

Possible

Solution

Verify the sensor offset and reactivate with the correct
settings if necessary.

Advertising