IETG FlowHawk Manual

Configuring the Monitor to Support the Ru-33

Configuring the FlowHawk to support the Telog Ru-33 requires
setting three parameters in Profile when establishing a new
location.

Scan frequency This represents the interval at which

the monitor takes readings. These readings are available
for retrieval directly from the monitor, but are overwritten
with new data at each new interval. Therefore, IETG
recommends using the same interval for both the scan rate
in the monitor and the rate at which the Telog unit requests
the data from the monitor.

Modbus ID This represents the Modbus slave ID that

will be used by the Ru-33 to indicate the specific monitor
from which to request the data. IETG recommends using
55 to identify the FlowHawk.

Modbus Delay Response This represents the amount

of time, in milliseconds, following the request before the
monitor will return the requested Modbus output data to
the Ru-33. IETG recommends maintaining the 10 ms
default setting.

Connecting the Monitor to the Ru-33

Connecting the FlowHawk to the Ru-33 requires the Triton-Telog
Comm Cable (p/n 8000-0054-01). First, connect the black, metal
connector (with gray, plastic boot) from the comm cable to the
COMM port on top of the FlowHawk. Then, connect the black,
plastic connector from the comm cable to the sensor port on the Ru-
33. The sensor port is the 9-hole port to the right of the antenna
connector and to left of the small 5-pin connector.