Update firmware via xmodem, System parameters – ADS Environmental Services IETG FlowHawk QR 775012 A4 User Manual

Page 374

Advertising
background image

D-6

IETG FlowHawk Manual

monitor is not detected, a Failed to detect monitor message will
display.

Update Firmware via XMODEM

Selecting this submenu enables you to update the EMUX firmware
through the terminal emulator or another serial connection that
supports XMODEM.

Note: Currently, EMUX firmware updates cannot be
performed through wireless communication. Therefore,
perform all firmware updates using a serial connection.

From the appropriate screen on the terminal emulator, enter or
browse to the desired file for the update (.BIW) and select Xmodem
as the protocol. Then, initiate the update through the emulator.
Once the file transfer is complete, the EMUX will restart
automatically.

Enter three bangs (!!!) to return to the onboard diagnostics main
menu.

System Parameters

Selecting this submenu displays the system parameters configured
for the EMUX. These parameters can be updated only by activating
a monitor attached to the EMUX through Profile

®

. The parameters

displayed should accurately reflect the configuration of monitor
communications with the EMUX.

Following is a brief description of each parameter:

Modbus ID This represents the ID used by the RTU to

indicate the specific EMUX/monitor from which to request the
data.

Modbus response delay This represents the amount of

time, in milliseconds, following a request before the monitor
returns the requested data to the RTU through the EMUX.

Scan frequency This represents the interval at which the

EMUX requests the latest data from the monitor.