6-52 IETG FlowHawk Manual

Modify the parameters as necessary, and then select OK once
complete.

DAC Gain Choose the method for managing the

amplification of the analog signal that returns to the sensor:

Auto Select this option to allow the monitor to

automatically optimize the amplification of the signal.

Manual Select this option to enter a fixed value

between -12db (lower amplification) and +12db (high
amplification) at which to amplify the signal.

Surcharge Detection Select the Disable checkbox to

prevent the monitor from implementing the Surcharge
Detection function when processing smart depth data. For
locations that typically do not experience surcharges,
disabling Surcharge Detection eliminates the processing
time required for this function and, as a result, conserves
monitor power.

Sensor

Physical Offset Enter the physical distance from the face of
the sensor to the crown (top) of the pipe. Refer to Chapter 3,
Sensor Installation and Connection, for instructions on
measuring the physical offset.
The default offset is 32 mm.

Pipe Height This field displays the pipe height at the monitor
location, based on the installation table. This parameter is not
editable from this location.

Channel Select the channel number of the port on top of the
monitor to which this device is connected.