8-58 IETG FlowHawk Manual

Problem

Individual PEAKVEL_1 or PEAKVEL_3 values
consistently read -9.99.

Possible

Causes

Sensor may be faulty.

5-volt fuse on regulator may have blown.

Possible

Solutions

Replace the Peak, Slimline Peak, or Surface Combo
Sensor.

Replace the 5-volt blue fuse (when replaceable) on the
regulator or return the monitor to IETG for regulator
replacement.

Caution: Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5-volt (blue)
replacement fuse on the regulator in the monitor.

Surface Velocity Subsystem

The following tables contain general troubleshooting techniques for
the surface velocity subsystem corresponding to the Surface Combo
Sensor.

Problem

Velocity readings are erratic or incorrect.

Possible

Causes

Sensor may be dirty or covered by debris.

Maximum Velocity setting may require adjustment.

Possible

Solutions

Clean the sensor transmit/receive surfaces.

Fine-tune Maximum Velocity setting in device
parameters.