Emergency stop connections, Emergency stop connections 129, Emergency stop output – MTS Model FlexTest SE User Manual

Page 129: Emergency stop input, Cable specifications

Advertising
background image

Emergency Stop Connections

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

129

Emergency Stop Connections

Connectors J23 E-STOP Out and J24 E-STOP In are located on the Model
493.73 HPU transition board. The Emergency Stop switch and signal are part of
a controller-wide interlock system.

Emergency

stop output

Connector J23 E-STOP Out provides an output to external devices when an
emergency stop signal is generated. The inactive status is shown in the figure.

Emergency

stop input

Connector J24 E-STOP In accommodates an external emergency stop switch.
As shown, several connectors maintain the continuity of the emergency stop
interlock.

Cable specifications

J23 E-STOP Out is a 9 pin type D male connector

9-contact type D female EMI connector

Cable—24 AWG, 10 conductor with overall foil shield, (Carol C0745 or
equivalent) with drain wire connected to metallized plastic backshell to the
chassis.

1

2

3

Emergency

Stop

J23

To

External

Devices

From

Chassis

+24 V

13

8

7

5

E-Stop

Signal

J24

E-Stop

Relay

External

E-Stop

Advertising