MTS Model FlexTest SE User Manual

Page 98

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Full-Range Digital Universal Conditioner (FRDUC)

FlexTest GT Controller Connections

98

Excitation sense

Jumpers W5 and W7 select the local (on board) or remote (through a cable)
excitation sensor connection.

The local selection is used with 4-wire transducer connections.

The remote selection is used with 8-wire transducer connections.

Jumper W7 configures the + excitation signal.

Jumper W5 configures the – excitation signal.

Both jumpers should be set for the same configuration.

±Shunt cal

The shunt calibration resistor can be installed on the Full Range Digital
Universal Conditioner module, or, in the transducer ID module. R33 assigns a
positive or negative excitation to the shunt calibration resistor if installed on the
Full Range Digital Universal Conditioner module (not the transducer ID
module).

Single ended

excitation

Jumper W8 connects -EX to ground for situations requiring single- ended
excitation.

W7
W5

W
W

W5

W7

W5

W7

Local

Remote

Default (No Resistor)
+EX Shunt

-EX Shunt

R33

Default

–EX = –EX

–EX = GND

W8

W8