D/a connections – MTS Model FlexTest SE User Manual

Page 117

Advertising
background image

Analog I/O Connections

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

117

D/A Connections

Digital-to-analog mezzanine cards support up to six program or readout signals
to external devices. Each D/A signal is an analog output within ±10 volts.

Each readout signal is from a 16 bit digital to analog converter.

Each output is a ±10 V analog output.

The Model 493.46 D/A Mezzanine Card provides analog signals to the
Model 493.76 Analog Out transition module. Analog outputs are available
at the BNC connectors of a Model 493.76 Analog Out transition module for
external devices (DVM, oscilloscope).

+

-

+

-

+

-

Output 1

Output 2

Output 3

Output 4

Output 5

Output 6

+

-

+

-

+

-

1
2
3
4
5
6
7
8
9

10

11

12
13
14
15

J4 - J7

From

D/A

To

External

Device

Advertising
This manual is related to the following products: